B&K Reference 7250 Series II Manual de Instruções Página 31

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 108
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 30
31
Outputs
Current output / Loop current
The loop current is controlled by the process variable selected
in the configuration.
The current start and end can be set to represent any desired
value. To check connected peripherals (e.g. limit switches,
controllers), the loop current can be manually specified
(see Pg. 41).
HART communication
The Transmitter Cond Ind 7100e/2(X) can be remote-controlled
via HART communication. It can be configured using a hand-
held terminal or from the control room. Measured values, mes-
sages and device identification can be downloaded at any time.
This allows easy integration also in fully automatic process
cycles.
A list of the HART commands can be found in the “Transmitter
Cond Ind 7100e/2(X) Transmitter-Specific Command
Specification”: www.mtpro.com/transmitters.
Alarm
The alarm delay is configurable.
Error messages can also be signaled by a 22 mA loop current
(see Configuration, Pg 55).
The alarm LED on the front panel can be configured as follows:
HOLD off: Alarm: LED flashing
HOLD on: Alarm: LED on. HOLD: LED flashing.
70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 31
Vista de página 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 107 108

Comentários a estes Manuais

Sem comentários