Transmitter Cond Ind 7100e/2(X)
Calibration
Calibration adjusts the device to the sensor.
Activate with cal
Enter passcode*:
Entry of cell factor 1100
With calibration solution 0110
Product calibration 1105
Zero point 1001
Temp probe adjustment 1015
Select with key, edit number with
key, proceed with enter key
(End with cal, then enter.)
The loop current is frozen (at its last
value or at a preset fixed value,
depending on the configuration),
Sensoface is off, mode indicator
“Calibration” is on.
Activate
Hold
Input errors
The calibration parameters are
checked during the input. In the case
of an incorrect input ”Err” is dis-
played for approx. 3 sec. The incor-
rect parameters cannot be stored.
Input must be repeated.
End
End with cal.
Safety prompt:
The measured value and Hold are dis-
played alternately, “enter” flashes.
Press enter to end the Hold mode.
The measured value is displayed.
The output current remains frozen for
another 20 sec (HOLD icon on,
“hourglass” flashes).
HOLD icon
During calibra-
tion the transmit-
ter remains in
the Hold mode.
* Factory setting, for passcode editing, see Pg 56
70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 58