CHAPTER 11
Logging (Optional Module)
101
11.3.1 The Profile View
The Profile view displays a profile of a logged broadband parameter (dB versus time). This is
very convenient when marking sound categories on-line or annotating the measurement. See
Fig.11.2.
Fig.11.2
Profile View (showing
extended status field)
Status Field
The Status Field has been extended to include an extra line of information below the two
existing lines of information covered previously in the Sound Level Meter and Frequency
Analyzer (see Fig.11.2).
This extra line of information allows you to:
• Select whether the results of the Total measurement from the Logged measurement, or from
the Logged(100 ms) measurement are displayed. Select Total to display the measurement
parameters of the Total measurement in all the views – the Broadband and Spectrum views
will then display parameters or spectra similar to the Sound Level Meter or Frequency
Analyzer. (The Profile View will be empty, because the Total measurement contains only a
single set of parameters). Select Logged to display the measurement parameters from the
logging intervals. The cursor in the profile selects which logging interval is displayed in all
the views. Select Logged(100 ms) to display the L
Aeq
, L
AF
or L
AS
in the profile from the
100 ms logging intervals. This setting does not display spectra or parameters in other views
• View the start time of the measurement (for Total) or the start time of the current logging
interval (for Logged or Logged(100 ms) if measuring and the profile is not frozen) or the
start time of the logging interval pointed out by the cursor. Tap on the start time in any of
the views to select data from another logging interval