When the Modem parameter is set to GPRS/EDGE/HSPA Dialup, then a connection is made
using the settings in Dialup Networking. The Status parameter reflects the current status of the
network connection – e.g., ‘Connected’ or ‘Disconnected’. If the connection fails, then the
analyzer will try to reconnect – until it succeeds, or the Modem parameter is set to Disabled.
If the connection succeeds, then the instrument will automatically get an IP address, which is
readable (but not accessible) in Network Settings – for more details see section 8.3.
The IP address is the address on the WAN (Wide Area Network) or Internet. Use this address in
BZ-5503 or an Internet browser to connect to the analyzer.
Note: In some cases you might be able to get a static IP address from the Network Operator,
which means that the analyzer will always get the same IP address, when connected to the
network. However, typically the IP address will be renewed each time the analyzer connects to
the network. In order to deal with this in BZ-5503 or an Internet browser you need to use
DynDNS – see section 8.2.3.
Modems with USB Interface
Note: Turn off the analyzer when connecting the analyzer to a modem.
The following modems have been tested:
Airlink Fastrack Xtend GPRS, EDGE and HSPA from Sierra Wireless
(www.sierrawireless.com)
Modems with an USB interface can be connected to the USB Type A socket on hardware
version 4.
Modems with RS232 Serial Interfaces
The following modems have been tested:
FASTRACK Supreme 10 and 20 GSM/GPRS/EDGE Modem from Wavecom
Airlink Fastrack Xtend GPRS, EDGE and HSPA from Sierra Wireless
(www.sierrawireless.com)
For hardware version 4, modems with an RS232 serial interface can be connected using a
USB to serial converter inserted in the USB Type A socket of the analyzer. The following USB
to serial converters have been tested:
USB to RS232 Converter UL-0250 (see “Ordering Information” on page 136.)
ES-U-1001-R10, ES-U-1001-B10 and ES-U-1001-A from EasySync
(www.easysync-ltd.com)
USB to RS232 Standard, USB to RS232 Professional and USB to RS232 Mini STD from
U.S. Converters (www.usconverters.com)
For hardware versions 1 3, use a compact flash to serial converter inserted in the compact
flash slot of the analyzer. The following compact flash-to-serial converter has been tested:
Serial I/O CF Card – Ruggedized from Socket (www.socketmobile.com)