B&K 2250 Light Manual do Utilizador Página 131

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 130
CHAPTER 13
Tone Assessment Option BZ-7231 – 1/3-octave Method
123
Fig.13.6
The differences to the left
and right of the selected
frequency band are
shown in the tone
parameter panel – in this
example, there is a
11.4 dB difference to the
left and a 3.9 dB
difference to the right
The information shown in the value panel can be changed by tapping in the field. From the tone
parameters, the adjustment and the selected tone standard can be selected. See Fig.13.7.
Results from the tone assessment can also be found by tapping on the Broadband tab and
browsing the Broadband view in a similar way.