B&K 50 Series II Manual do Utilizador Página 61

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 60
47
4. Connect the external analog control pins from the rear panel of the instrument to your
voltage control source as shown below:
Figure 13 - Analog Control Setup
5. Set the instrument to CC mode and enable the input to begin controlling with the
external voltage source.
Do not connect more than 30 V into the external program terminal pins or damages may
occur.
External Input On/Off Control
The electronic load’s input can be enabled (ON) or disabled (OFF) via the external control pins
labeled ON and ground pin ( ).
Enable (ON) input: Short the two pins together, as shown below:
Rear Panel Terminals
Rear Panel Terminals
0 10 V
External Voltage Source
Vista de página 60
1 2 ... 56 57 58 59 60 61 62 63 64 65 66 ... 101 102

Comentários a estes Manuais

Sem comentários