65
0 ≤ T
pf
≤ (T
on
+ T
off
)
Tpf = P/F Delay Time
T
on
= On Time
T
off
= Off Time
Use the numeric keypad or rotary knob to enter a value. Then press .
8. If there are more than one active sequence, the load will repeat the prompts to
configure the On Time, Off Time, and P/F Delay Time. Follow steps 7 9 for each
sequence.
9. Once all sequences’ parameters are configured, the display will show:
Select Complete-Stop or Failure-Stop, then press .
Complete-Stop: Stop test when program is complete.
Failure-Stop: Stop test when testing fails.
10. The load will prompt to enter a value for Chain Program File. This specifies the next
program to run immediately after the end of the test. Select between 0 10. Select 0 if
no other programs will be run at the end of the test. Otherwise, select the program
number to run after the test is completed. Use the numeric keypad or rotary knob to
enter a value and then press to confirm.
11. Lastly, the load will prompt to select a number to Save Program File. This is the
program number that will be referred to when recalling all the configured settings from
previous steps or chaining with another program. Select between 1 10 with the
PROGRAM
CompleteStop FailureStop