Hand-held Analyzer Type 2250 –
Instruction Manual
44
Response to tone bursts for exponential averaging detectors according to IEC 60651 and DIN 45657:
Note: The maximum response to test tone burst referred to response to continuous signal is calculated from the formula in IEC 60651
appendix C.
4.10.2 Linear Averaging
Linear averaging times: 1 second to 24 hours, in steps of 1 second
Settling time according to IEC 60804: < 1 second
Nominal delay time between operation of the reset facility and re-initiation of a measurement
according to IEC 61672–1: < 2 seconds
Time interval after completion of a measurement before a reading is displayed according to
IEC 61672–1: < 1 s
Minimum hold time according to IEC 60804: Results of a Timed Measurement are held until a new
measurement is initiated or the result is reset.
4.10.3 Peak
Peak onset time according to IEC 60651: <100µs
Table 4.9
Response to tone
bursts for exponen-
tial averaging
detectors
Time
Weighting
Duration of
test tone burst
(ms)
Maximum
response to test
tone burst
referred to
response to
continuous signal
(dB)
Standard
tolerances on
maximum response
(dB)
Type 2250
tolerances on
maximum
response
(dB)
Continuous 0
F 200 –0.98 ±1 ±0.1
100 –2.59 ±2 ±0.1
50 –4.82 ±2 ±0.1
20 –8.30 ±2 ±0.1
10 –11.14 ±2 ±0.1
5 –14.07 ±2 ±0.1
2 –17.99 ±2 ±0.1
1 –20.99 ±2 ±0.1
0.5 –23.99 ±2 ±0.1
0.25 –26.99 ±2 ±0.1
S 2000 –0.63 – ±0.1
500 –4.05 ±1 ±0.1
200 –7.42 – ±0.1
50 –13.12 – ±0.1
I 20 –3.61 ±1.5 ±0.1
5 –8.76 ±2 ±0.1
2 –12.55 ±2 ±0.1
BE171211.book Page 44 Thursday, May 27, 2004 1:01 PM