CHAPTER 4
Specifications
49
4.13 Electrical Input to Connector Panel of Type 2250
4.13.1 Input Socket (Rear Socket)
Triaxial Lemo socket used for Direct input as well as CCLD input.
Direct Input:
Max. input level: ±14.14 V
Peak
, no damage for signals up to ±20 V
Peak
Input impedance: > 1 MΩ
CCLD Input:
Max. input level: ±7.07 V
Peak
(no indication for violation of this level), no damage for signals in the
range –10 to +25 V
CCLD current: 4 mA @ 25°C
4.13.2 Trigger Socket
Triaxial Lemo socket:
Input Range: ±20 V
Peak
, no damage for signals up to ±50 V
Peak
Supply for Remote Control Switch: 5 V through Silicium diode and 5 k resistor
Input impedance: > 1 MΩ
4.14 Electrical Output from Type 2250
4.14.1 Output Socket
Triaxial Lemo socket:
Output Signal: Input conditioned; A-, C-, or Z-weighted; covers the nominal range 20 to 140 dB
Gain Adjustment: –80 to +60dB
Max. peak output level: ±4.46 V
Output impedance: 50 Ω
Min. load impedance: 15 kΩ ⎟⎟ 1 nF for < 0.2 dB attenuation from DC to 20 kHz, short-circuit
proof without affecting the measurement results
Max DC offset: ±15 mV
4.14.2 Headphones Socket
Stereo MiniJack socket:
Output Signal: Input signal can be monitored using this socket with headphones/earphones
Gain Adjustment: –80 to +60dB
Max. peak output voltage: ±1.4 V
Output impedance: 2.2 Ω in each channel, short-circuit proof without affecting the measurement
results
BE171211.book Page 49 Thursday, May 27, 2004 1:01 PM