154
Hand-held Analyzer Types 2250 and 2270 – User Manual
Fig.14.5
Selecting the bands for
serial measurements
Left: Selecting the band
for serial measurement
Right: The arrow is
pointing to the “s” symbol,
which indicates the
selected frequency band
Note: A small s appears above the selected frequency bands (Fig.14.5).
4) When you have selected the frequency bands you want to re-measure, press the
Start/Pause pushbutton.
The analyzer will then start at the lowest selected frequency band, generate noise at this band, make
the measurement and automatically step to the next selected band. This will continue until all selected
bands have been measured.
5) Press the Save pushbutton to save and merge the bands with the measurement you have
already done (be sure to select the right position before saving).
Combine Parallel and Serial Measurements in an Automated Sequence
You can speed up the measurement process if you start by investigating whether you need to
use serial measurements at certain frequency bands, or not. If you need to measure some bands
serially, then you can select Measurement Control from the Setup screen, and set the
Measurement Mode to Combined. Then select the bands you want to measure serially (see
above).
Press the Start/Pause pushbutton to make a normal parallel measurement and then do the serial
measurements – all in an automated sequence.
14.2.6 Annotating Projects and Positions
While using the Reverberation Time Software you can annotate the Reverberation Time project using
the normal method of adding annotations to a project. The annotations can then be viewed by tapping
on the Annotations icon or tapping the Main Menu icon and selecting Explorer from the list of
options. See section 3.5.
However, your project can contain a lot of measurements, one per position, so you can annotate
any position in the project. Signal recordings are automatically attached as annotations to the
positions. Annotations at positions are managed in the Overview display – described in the
next section.