APPENDIX A
Setup Parameters
247
Criterion Level 0 to 140 dB Criterion Level is the sound level allowed for
an 8-hour period and would yield a 100%
dose.
The criterion level must be set according to
the legislation with which you will have to
comply.
Used for calculation of Dose, ProjDose,
DoseUQ, ProjDoseUQ.
U = Time Weighting for Lav: F or S
Q = Exchange Rate for Lav: 4, 5 or 6 dB
PeaksOver Level 0 to 200 dB Any peak levels that exceed the level set here
will be counted.
Used for calculation of #XPeaks(>NNNdB)
X = A, B, C, Z or G, set by Setup > Frequency
Settings > BB Peak, NNN is the PeaksOver
Level
Note: Two other peak counters are preset to
count peaks over 135 dB and 137 dB,
respectively
Exchange Rate for Lav 4 dB
5dB
6dB
The increase in noise level that corresponds
to a doubling of the noise level is determined
by Exchange Rate for Lav.
Set Exchange Rate for Lav as required by
your local standards.
Used for calculation of LavUQ, TWA, TWA
v
DoseUQ, ProjDoseUQ.
U = Time Weighting for Lav: F or S
Q = Exchange Rate for Lav: 4, 5 or 6 dB
Note: TWA and TWA
v
requires U = S and
Q = 5
Time Weighting for Lav F
S
Lav will be time weighted in accordance with
this parameter.
Used for calculation of LavUQ, TWA, TWA
v
DoseUQ, ProjDoseUQ.
U = Time Weighting for Lav: F or S
Q = Exchange Rate for Lav: 4, 5 or 6 dB
Note: TWA and TWA
v
requires U = S
and Q = 5
*. Not available for Reverberation Time Software BZ-7227.
Table A.16 (Cont.) Occupational Health Parameters
*
Parameter Values Comment