Hand-held Analyzer Types 2250 and 2270 – User Manual
If you have selected 1/1-octave, an Accelerometer as transducer, or a Direct input, tone
assessment will be performed, but a smiley will be displayed. When you tap the smiley it will
advise you to use a Microphone.
For an overview of all the remedies associated to the various smileys, please refer to Table 15.2
Table 15.2
Overview of smileys and associated remedies
Note: The ISO method does not specify the weighting filter to be used. The A-weighting filter
is selected by default, but all selectable broadband filters may be applied without warnings.
15.9 Recalling Saved Measurements
Saved measurements using the Frequency Analyzer, Logging and Enhanced Logging
templates can be opened and the Tone Assessment can be applied. Please refer to section 3.4
and section 6.3 for information on how to save and recall measurements, respectively.
Smiley Colour Explanation Remedy – Tip for Solution
Red No tone assessment on 1/1
octaves
There is no method for assessing tones when
the spectrum is analysed in 1/1-octave bands,
select 1/3-octave instead
Red Analysis on ‘Accelerometer’ The analysis is carried out on an accelerometer
input, select a microphone
Red Spectrum must be L
ZFmin
When Italian Law is selected the analysed
spectrum must be L
ZFmin
Red No tone assessment on inst.
parameters
It is not possible to make tone assessments on
instantaneous parameters (‘Fast’ or ‘Slow’)
Yellow Analysis on ‘Direct’ The analysis is carried out on a ‘Direct’ input,
select a microphone
Yellow Setup different from ISO
standard
The ISO is selected and the measurement setup
deviates from the proposals in the ISO standard
Yellow Min or Max spectrum When the ISO standard is selected the analyzed
spectrum should be L
eq
Yellow Tone freq outside proposed
ISO range
In the ISO standard the defined frequency range
for tone assessment is from 25 Hz to 10 kHz. A
tone is detected outside this frequency range
Yellow Tone freq out of proposed DM
range
In the standard according to Italian Law the tone
assessment must be performed in the range
from 20 Hz to 20 kHz. A tone is detected outside
this frequency range