186
Hand-held Analyzer Types 2250 and 2270 – User Manual
16.6 Setting Up 2-channel Logging and Enhanced Logging
1) Select:
– The 2-Ch. Logging SLM project template for logging broadband data only from the
two inputs
– The 2-Ch. Logging project template for logging both spectra and broadband data from
the two inputs
– The 2-Ch. Enh. Logging project template for logging both spectra and broadband data
from the two inputs, making reports and periodically saving the data on the memory
card.
Note: 2-Ch. Enh. Logging is not available for hardware versions 1–3.
The project template is displayed at the top of the screen, if it does not display the correct
template tap the Project Template Bar at the top of the screen and select the correct
template from the drop-down menu. (See section 3.3.1 for more details on templates.)
2) Tap the Main Menu icon .
3) Select
– Setup > Frequency Settings to set Broadband frequency weighting parameters,
as required
– Setup > Statistics to set statistics parameters, as required
4) Set the parameters you want to log every 100 ms in Logged Broadband (100 ms).
You can specify up to 4 parameters from the two channels.
5) Set the broadband parameters you want to log in Logging Broadband.
You can either log all parameters from both channels or select up to 10 parameters from
the two channels. The fewer parameters you select the less space will be needed for saving
on the memory card.
6) Set the spectrum parameters you want to log in Logged Spectrum.
You can either log all spectrum parameters from both channels or select up to four
spectrum parameters from the two channels.
7) Set the Level Trigger parameter(s) as required.
You can level trigger on a parameter from either Ch.1 or Ch.2.
8) Tap to exit Setup.
Note: You should select the transducers before selecting the parameters to log and selecting the
level trigger parameter(s). Changing the transducer afterwards might correct the selected
parameter; e.g., if you have selected to log Ch.1 LAeq and change the transducer at channel 1
from a microphone to an accelerometer, then the Ch.1 LAeq will not be available anymore and
the parameter will be set to Off.