APPENDIX A
Setup Parameters
243
Automatic Gain Control
On
Off
To ease identification of sound sources, the gain can
be automatically adjusted to keep the average level
within a 40 dB range. When playing back the recorded
signal, you will then hear clearly the whole signal
content, whether the level has been 20 dB or 140 dB.
Set Automatic Gain Control to On to convert the
recorded signal. The recorded signal is stored as a
16-bit .WAV file.
Set Automatic Gain Control to Off for recording the
signal with a fixed gain – then set Resolution = 24 bit
(recommended) to cover the full 120 dB dynamic
range (from max. input level and down) or select
Resolution = 16 bit and specify the Peak Recording
Level to fit the signal.
Note: If the sound contains very high levels at low
frequency, then a fixed gain is recommended
Resolution 24 bit
16 bit
Set Resolution to 24 bit to cover the full dynamic
range. Set Resolution to 16 bit to cover up to 96 dB
from Peak Recording Level and down.
Note: You will only be warned about overload in the
recorded signal if Automatic Gain Control is On or if
Resolution is 24 bit or if Peak Recording Level is set to
its maximum value
Note: 24-bit recording not possible in BZ-7227
Peak Recording Level 145 dB
135 dB
125 dB
115 dB
105 dB
95 dB
85 dB
75 dB
For Automatic Gain Control = Off and Resolution = 16
bit the recorded wave file has a dynamic range of up to
96 dB. When playing back on the analyzer, the
dynamic range of the output is approx. 75 dB. When
playing back on a PC it might be even lower. Set Peak
Recording Level to fit the signal.
The values for Peak Recording Level take the
sensitivity of the attached transducer into account. The
values shown in the list here are nominal values for a
Type 4189 or 4190 microphone.
Hint: Monitor the L
peak
value during a trial
measurement before selecting the Peak Recording
Level
Note: For recording in 2-channel templates the Peak
Recording Level will automatically be set in
accordance with the Input > Range Setting parameter
(i.e., High Range or Low Range)
Table A.13 (Cont.) Signal Recording parameters
*
Parameter Values Comment