20
Hand-held Analyzer Types 2250 and 2270 – User Manual
7) Add any spoken comments to the measurements by pressing the Commentary pushbutton,
and add any written comments by tapping the Main Menu icon and selecting Add
Note to Current Measurement from the drop-down that appears.
8) To view and organise your data, tap the Main Menu icon and select Explorer.
Note: You are not required to set any measurement ranges, the analyzer has a dynamic range of
more than 120 dB, from 140 dB down to the noise floor of the microphone, (if the microphone
has nominal sensitivity).
3.2.1 Congratulations!
You should now be familiar with the basic principles of the analyzer. If you need more help, the
following section goes into the measurement process in more detail. If not, please refer to
Chapter 1.
3.3 Making a Measurement
3.3.1 What is a Project Template?
A project template contains all the common display settings and measurement setups required
to perform a noise measurement. The template does not contain any measurement data – these
data are saved as individual projects, stored in job folders, see section 6.1.1. The project
templates covered by this manual, are:
Sound Level Meter (included in BZ-7222 software)
Frequency Analysis (included in BZ-7223 software)
Logging (included in BZ-7224 software)
Enhanced Logging (included in BZ-7225 software)
Reverberation Time (included in BZ-7227 software)
Note 1: Signal Recording Option BZ-7226 does not contain a specific template signal
recording options are available in all templates.
Note 2: Tone Assessment Option BZ-7231 does not contain a specific template tone
assessment options are available in all templates included in BZ-7223/24/25/30.
Note 3: Low Frequency Option BZ-7234 works with all the templates included in
BZ-7222/23/24/25
Note 4: 2-channel Option BZ-7229 (for Type 2270) enables 2-channel operation for all
2-channel templates included in BZ-7222/23/24/25/28 and for Signal Recording Option
BZ-7226 and Low Frequency Option BZ-7234.
If you make any changes to the settings in a project template, an ‘
*
’ will appear next to the
template name to indicate that the new settings have not been saved. Tap the template name at
the top of the display to open the Template Explorer (or select it from the Main Menu )
and tap the Save icon to save the settings in the current template.