Linear Operating Range: In accordance with
IEC 61260, 1/3-octave: 20.5 to 140 dB
SAMPLING FOR OCTAVE OR 1/3-OCTAVE
STATISTICS
X = frequency weightings A or B
The Statistics can be based on either L
XF
or L
XS
:
•Statistics L
XFN1-7
or L
XSN1-7
are based on sampling
L
XF
or L
XS
, respectively, every T ms into 1 dB wide
classes over 150 dB;
T = 100 for frequency range set to 12.5 – 20 kHz
T = 200 for frequency range set to 6.3 – 20 kHz
Full distribution can be saved with measurement
Displays
MEASUREMENT DISPLAYS
Spectrum: One or two spectra superimposed + A/B
and C/Z broadband bars
Table: One or two spectra in tabular form
Y-axis: Range: 5, 10, 20, 40, 60, 80, 100, 120, 140 or
160 dB. Auto zoom or auto scale available. Sound
data are displayed as dB values, vibration data as dB
values or in physical units (SI units (m/s2) or US/UK
units (g)), direct data as voltage in dB or V.
Cursor: Readout of selected band
Generator
INTERNAL GENERATOR
Built-in pseudo-random noise generator
Spectrum: Selectable between Pink and White
Crest Factor:
Pink Noise: 4.4 (13 dB)
White Noise: 3.6 (11 dB)
Bandwidth: Selectable:
• Lower Limit: 50 Hz (1/3-oct.) or 63 Hz (oct.)
• Upper Limit: 10 kHz (1/3-oct.) or 8 kHz (oct.)
Output Level: Independent of bandwidth
• Max.: 1 Vrms (0 dB)
• Gain Adjustment: –60 to 0 dB
When bandwidth is changed, the level for all bands is
automatically adjusted to comply with the set output
level
Correction Filters: For sound sources Type 4292,
Type 4295 and Type 4296: Flat or Optimum
Repetition Period: 175 s
Output Connector: Output socket
EXTERNAL GENERATOR
Selectable as alternative to Internal Generator (for
microphone input only)
To control external noise generator, set:
• Levels: 0 V (Generator off), 4.5 V (Generator on)
• Rise-time and Fall-time: 10 s
The noise generator is turned on and off automatically
during the measurement
Escape Time: 0 to 60 s
Build-up Time: 1 to 10 s
The generator can be turned on and off manually for
checking equipment and sound levels
Software Specifications – Logging Software BZ-7224
The specifications for BZ-7224 include the specifi-
cations for Sound Level Meter Software BZ-7222. BZ-
7224 adds:
Logging
MEASUREMENTS
Logging: Measurement data logged at pre-set periods
into files on:
• SD Card: All hardware versions
• CF Card: Hardware versions 1 – 3
• USB Memory Stick: Hardware version 4
Logging Period: From 1 s to 24 hr with 1 s resolution
Fast Logging: L
AF
, L
AS
and L
Aeq
(sound input) or Fast
Inst, Slow Inst and Linear (vibration and Direct input)
can be logged every 100 ms, irrespective of logging
period
Broadband Data Stored at each Logging Interval:
All, or up to 10 selectable broadband sound data incl.
Trig. Input Voltage, Weather data, and L
Aeq,T,mov
(for
vibration and direct input: up to 5 parameters)
Broadband Statistics Stored at each Logging
Interval: Full distribution, or none (sound input only)
Spectrum Data Stored at each Logging Interval: All,
or up to three selectable spectra (license for BZ-7223
required)
Spectral Statistics Stored at each Logging Interval:
Full distribution, or none (sound input only, license for
BZ-7223 required)
Logging Time: From 1 s to 31 days with 1 s resolution
Measurement Total: For the logging time, in parallel
with logging: All broadband data, statistics and spectra
(license for BZ-7223 required)
MARKERS
One data exclusion marker and four user-definable
markers for on-line marking of sound categories heard
during the measurement
Events can be set manually