
CHAPTER 15
Tone Assessment Option BZ-7231 – 1/3-octave Method
173
Italian Law
When tone assessment according to Italian Law is selected, the analyzed spectrum must be the
Z-weighted minimum level measured in each 1/3-octave band, L
ZFmin
. The L
ZFmin
parameter
is automatically selected on the measurement page when you select DM 16-03-1998.
The test for loudness contours is set by default to the 1987 Free-field option of the ISO 226
contour parameters. You can also select the 1987 Diffuse-field and 2003 Free-field options
from the ISO 226 contour parameters, if required. See Fig.15.6.
Fig.15.6
Setting the loudness
contour options
15.4.3 Setting up the Measurement using the Default Setup
Select the Tone parameter above the spectrum and start the measurement by pressing the
Start/Pause pushbutton . If the selected setup parameters do not comply with the selected
tone assessment method, the following pop-up window will be displayed:
Fig.15.7
Pop-up window for
tone measurement
setup check
Comentários a estes Manuais