APPENDIX A
Setup Parameters
229
A.5 Measurement Control – for Reverberation Time
Software BZ-7227
Table A.5 Measurement Control parameters – for Reverberation Time Software BZ-7227
Parameter Values Comment
Map Based Measurement No
Yes
If Map Based Measurement = Yes you can define
the source positions and receiver positions
graphically on a map – otherwise the
measurements are just numbered from position
number one (Pos. 1) onwards
Meas. All Pos. for Each Source No
Yes
Parameter only available if Map Based
Measurement = Yes.
Set to Yes for measuring all positions for each
source position.
Set to No for measuring a specific number of
receiver positions per source position
No. of Positions per Source 1 to 10 Parameter only available if Meas. All Pos. for Each
Source = No
Set number of receiver positions you want to
measure per source
Increment Sources First
Receivers First
Manual
Parameter only available if Map Based
Measurement = Yes.
Allows you to select the quickest method of
measurement, when positioning your analyzer in
relation to source and receiver positions. (Typically,
Sources First is best for Impulsive Excitation and
Receivers First is best for Interrupted Noise
excitation)
Automatic Save No
Yes
Set to Yes to automatically save the decay after
each measurement
Max. Decay Time 1 to 30 s Sets the maximum duration of the decay
measurement. If the decay finishes in a shorter
time, decay measurement stops automatically
Note: If you set Max. Decay Time above 20 s, then
the measurement will not stop before this time has
elapsed