The pop-up window can be de-activated for the rest of the measurement session. To activate it
again, you will have to reload the template or restart your analyzer. This automatic check is
deactivated when the Tone Assessment parameter is set to Off in the Setup screen.
To reset all the relevant parameters to the default setup, tap the OK button. (The default setup
includes preset parameters that comply with the selected standard. These are described in Table
15.1.) To measure with the settings you set up manually, tap the Cancel button.
Note 1: Even if you have set the Tone Assessment parameter to On (Setup screen), the
calculation is not carried out until the Tone parameter is selected above the spectrum.
Note 2: You may find the Template Explorer (accessed from the Main Menu ) useful to
save your own measurement setup.
15.4.4 Signal Recording
If a valid license for the Signal Recording Option BZ-7226 is present, it is possible to record
the sound while measuring. (Refer to Chapter 13 for information about the Signal Recording
Option.)
Note: When analyzer recordings are to be used for re-analysis on a PC, be sure to select
Automatic Gain Control parameter to Off, under the Signal Recording parameters, and select
the Recording Quality parameter to High.
Table 15.1
Default parameters
used by the automatic
measurement setup
check
Setup Parameter Default Value
Selected standard: ISO 1996-2, Annex D
Adjustment 3 dB
Last band in low frequency range 125 Hz
Last band in middle frequency range 400 Hz
Level difference low 15 dB
Level difference middle 8 dB
Level difference high 5 dB
Selected standard: Italian Law (DM 16-03-1998)
ISO 226 1987 Free-field