238
Hand-held Analyzer Types 2250 and 2270 – User Manual
Level Trigger 1
†‡
Trigger On
Off
Set to On for setting Marker 3 (Level) when
the Trigger Parameter fulfils the Level Trigger
conditions (see below). The conditions are
checked every second.
Signal Recordings
c
can also be controlled by
these settings, see Setup > Signal
Recording > Recording Control
Start Time 00:00:00 to 23:59:59 Start Time defines when the level trigger is
enabled. Triggering is limited to the enabled
time
Stop Time 00:00:00 to 23:59:59 Stop Time defines when the level trigger is
disabled. Triggering is limited to the enabled
time
Hold Off 00:00:00 to 01:00:00 Hold Off specifies a time after a level trigger,
where level triggering is disabled. Can be
useful in reducing the number of triggers
Start Slope Rising
Falling
Set to Rising to start when level goes above
Start Level (and then stop when level goes
below Stop Level).
Set to Falling to start when level goes below
Start Level (and stop when level goes above
Stop Level)
Start Level –100 to 200 dB Start trigger conditions are fulfilled when level
crosses Start Level (in accordance with Start
Slope) for at least Start Duration seconds –
set Start Level here
Start Duration 0 to 420 s Start trigger conditions are fulfilled when level
crosses Start Level (in accordance with Start
Slope) for at least Start Duration seconds –
set Start Duration here. The Start Duration will
be limited by the internal recording buffer size
and the Recording Quality and Resolution
(Table A.18)
Stop Level –100 to 200 dB Stop trigger conditions are fulfilled when level
crosses Stop Level (in accordance with Start
Slope) for at least Stop Duration seconds – set
Stop Level here
Table A.11 (Cont.) Trigger parameters
*
Parameter Values Comment